High flow rate control panel for oxygen, with automatic change-over and available both with reset and automatic; calibrated at 9 bar at the outlet, it reaches a maximum flow rate of 150 mc / h in oxygen in the manual reset version, 135 mc / h in the fully automatic version; it is available with pressure switch; consisting of two double reduction lines, it guarantees continuous supply with high pressure stability; the outlet connector is for a Ø 22 mm welded pipe; the panel is completed by a steel plate and a white PETG cover that is light and practical to install.
N.2 high pressure reducers 451 with brass piston and pressure regulation with stainless steel threaded dowel.
N.2 low pressure reducers 451 in brass and pressure regulation with stainless steel threaded dowel.
N.4 overpressure relief valves incorporated in the brass reducers.
N.1 outlet with fitting for Ø22 mm welded pipe.
N.2 shut-off valves for High Pressure in brass.
N.2 Low Pressure shut-off valves for maintenance.
N.2 brass non-return valves.
N.2 High Pressure gauges, 0-315 bar scale, class 2.5.
N.2 low pressure gauges, scale 0-16 bar, class 2.5.
N.2 sintered bronze filters.
N.2 pressure switches.
Stainless steel springs.
EPDM PEROX membranes.
Seal seat in NYLON.
EPDM PEROX O-Ring.
NYLON gaskets.
Cover in white PETG.
Powder coated metal fixing plate.
| CODE | DESCRIPTION |
| HK080HF | MAINTENANCE EQ REDUCTION BLOCK ECO EVO H.F. |
| HK080R1 | MAINTENANCE EQ ECO EVO REDUCER BLOCK |
| CODE | GAS | Q max. | P1 max. | P2 | INVERSION | ALARM SIGNAL | INLET CONNECTION | OUTLET CONNECTION |
| HQ066OX09KAP | O2 | 135 Nm³/h | 200 bar | 9 bar | AUTOMATIC | PRESSURE SWITCH | UNI-2 | Ø 22 mm weld-on pipe |
| HQ066OX09KSP | O2 | 150 Nm³/h | 200 bar | 9 bar | SEMI-AUTOMATIC | PRESSURE SWITCH | UNI-2 | Ø 22 mm weld-on pipe |